How often do I need to run Fitting Position?

Fitting Position is the qualification routine for determining location of the operating sphere for the Master Probe with RDS-CAA. This location needs to be re-determined every time the RDS head is removed and reinstalled. Also, this can be done whenever the machined is homed.